Gen Xpert XVI

Product ID : #36
Publish Date : Tuesday, 12.April 2016 16:34
View Count : 1149
Print Send to Friends
you will send:
Gen Xpert XVI
  • Reload Reload
Letters are not case-sensitive
Send
Gen Xpert XVI

OPTICS
Dye detection limit < 1 nM

 

OPTICAL CHANNEL CHARACTERIZATION
Channel Excitation Emission Calibrated Reporter Dyes
1 375–405 420–480 CF 1
2 450–495 510–535 FAM
3 500–550 565–590 Alexa Fluor® 532
4 555–590 606–650 Texas Red®
5 630–650 665–685 Alexa Fluor® 647
6 630–650 >700 CF 6


REACTION SITE THERMAL CONTROLS
Solid state heater and forced-air cooling at each site
Reaction chamber thermistors calibrated to ± 1.0°C
using National Institute of Standards and Technology
(NIST)-traceable standards
Up to 16 independently-controlled reaction sites


CARTRIDGES
Single-use disposable cartridges
Polypropylene construction


PERFORMANCE PARAMETERS
Heating ramp rates (max.):
10°C/sec from 50°C to 95°C
Cooling ramp rates (max.):
2.5°C/sec from 95°C to 50°C
Temperature duration accuracy:
± 1.0 sec from programmed time
Temperature accuracy:
± 1.0°C from 60°C to 95°C
Melt curve programmable ramp rates:
0.01°C/sec to 1.0°C/sec